User Contibuted 7411B PROCEDURE ================================================================================ INSTRUMENT: Hewlett Packard 5381A Counter No Options. DATE: 22-Jul-88 AUTHOR: User Contributed REVISION: 0.1 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 23 NUMBER OF LINES: 83 CONFIGURATION: Wavetek 278A Fluke 6060B =============================================================================== STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 ASK- J 1.002 DRAW 5 78 7 78 83 25 67 75 1.003 HEAD Initial Preparation. 1.004 DISP Remove the 4 Phillips screws from the bottom of the 1.004 DISP unit. 1.004 DISP Turn the unit upright & remove the top cover. 1.005 SET LINE: ON [DRAW17,20,F] 1.006 SET GATE TIME: 1S [DRAW33,20,F] 1.007 SET ATTENUATOR: X1 [DRAW33,80,F] 1.008 SET OSCILLATOR EXT/INT: INT [DRAW17,10,B] 1.009 HEAD {Amplifier Balance.} 1.010 DISP Temporarilly remove the top cover. 1.011 278 1MH 30mV SI S L 1.012 DISP Connect an oscilloscope to TP6. [DRAW44,86,T] 1.013 EVAL {Is the 'scope's display a 50% duty cycle square wave?} 1.014 278 1MH 30mV SI S L 1.015 DISP Adjust A1R19 for a 50% duty cycle square wave on 'scope. 1.015 DISP [DRAW22,64,T] 2.001 DISP Replace the top cover. 2.002 HEAD {Attenuator Check.} 2.003 278 1MH 30mV SI S L 2.004 EVAL {Is the counter triggered & does the display read near 1 MHz?} 3.001 SET ATTENUATOR: X10 [DRAW33,80,F] 3.002 278 1MH 300mV SI S L 3.003 EVAL {Is the counter triggered & does the display read near 1 MHz?} 4.001 SET ATTENUATOR: X100 [DRAW33,80,F] 4.002 278 1MH 3V SI S L 4.003 EVAL {Is the counter triggered & does the display read near 1 MHz?} 5.001 HEAD {Time Base.} 5.002 SET ATTENUATOR: X10 [DRAW33,80,F] 5.003 DISP If the unit has not been on for 5 minutes yet, let sit 5.003 DISP until 5 minutes have elapsed. 5.004 MESS Over range is indicated by a dot above & to the left of 5.004 MESS the MSD of the display. 5.005 6060 10000KH .01U 300mV -D 5.006 6060 10MH 300mV S 5.007 DISP Adjust OSCILLATOR ADJ for a display of 10 MHz +/- 10 Hz. 5.007 DISP [DRAW17,25,B] 6.001 MESS 6.002 SET GATE TIME: 10S [DRAW33,20,F] 6.003 M278 B4 6.004 278 11H .1U 2% 300mV -D SI L 7.001 278 100H .1U .0005% 300mV -D SI L 8.001 278 1000H .1U .0005% 300mV -D SI L 9.001 278 10000H .1U .0005% 300mV -D SI L 10.001 278 100000H .1U .0005% 300mV -D SI L 11.001 278 900000H .1U .0005% 300mV -D SI L 12.001 SET GATE TIME: 1S [DRAW33,20,F] 12.002 278 .011KH .001U 2% 300mV -D SI L 13.001 278 .1KH .001U .0005% 300mV -D SI L 14.001 278 1KH .001U .0005% 300mV -D SI L 15.001 278 10KH .001U .0005% 300mV -D SI L 16.001 278 100KH .001U .0005% 300mV -D SI L 17.001 278 1000KH .001U .0005% 300mV -D SI L 18.001 278 9000KH .001U .0005% 300mV -D SI L 19.001 M278 B0  19.002 SET GATE TIME: .1S [DRAW33,20,F] 19.003 6060 110KH .01U .0001% 300mV -D 20.001 6060 1000KH .01U .0001% 300mV -D 21.001 6060 10000KH .01U .0001% 300mV -D 22.001 6060 80000KH .01U .0001% 500mV -D 23.001 DISP Re-assemble the unit. You are finished with calibration. END